Signum 1-2-3
Classification Application W
69
Classification
Application W
With the Classification application, you
can determine whether the weight of a
given sample lies within the limits of a
defined weight class.
Characteristics
Classification with 3 or 5 weight classes.
Configured in Setup under:
Appl:
W: 4.8.
Signum
®
3:
Appl 2:
class: 4.8.
Enter the upper class limits using the
keypad or by saving weight values from
a load on the platform
Enter the upper limits of weight classes
as absolute values or as a percentage
of deviation from the upper limit of
Class 1 Configured in the menu under:
Appl: W: 4.9.
Signum
®
3:
Appl 2:
class: 4.9.
Activate Info mode via the I key
Toggle the main display between
classification display and tolerances
limits by pressing the w key
Automatic printout configured in Setup
under:
Appl:
W: 4.10.
Signum
®
3:
Appl 2:
class: 4.10.
Automatic taring of container weight.
Configured in Setup under:
Appl:
W: 3.7.
Signum
®
3:
appl: a.tare: 3.7.
Automatic initialization when you
switch on the scale with most recently
saved application data. Configured in
Setup under:
Appl:
W: 3.8.
Signum
®
3:
appl: a.start: 3.8.
Signum
®
3 only:
You can assign different functions to
the c key for deleting applications.
When you clear applications, you can
delete either the data stored for all
applications or just the data stored for
the active application. Configured in
Setup under:
Appl: sel.cf: 3.24.
(Sel. CF function key c in
applications)
Tare function:
1) If you store a tare (weight value) by
pressing the ) key, you can later enter
a value manually. The value you enter is
added to the stored tare value. Setting:
menu code 3.25.1 (factory setting)
2) A value entered manually overwrites
a stored tare value (weight value).
If you enter a value manually, a tare
value (weight value) stored later
overwrites the manually entered value.
Setting: menu code 3.25.2
Configured in Setup under:
Signum
®
2:
Appl: tare. f:: 3.25.
Signum
®
3:
Appl:
W:: 3.25.
Restore factory default settings.
Configured in Setup under:
Appl:
W:: 9.1.
Signum
®
3:
appl: def.app: 9.1.
To use the Classification application, you
need to enter the delimiters that separate
one class from another.
Limits between the individual weight
classes are required for the classification.
The lower limit of Class 1 is defined by
the preset minimum load. The other
classes are configured by defining their
upper limits. There are two ways to enter
the delimiters for classes 1 through 3
(or 5):
By saving the weight value indicated:
Each upper limit value, with the
exception of the highest class, is entered
using the keypad or by saving the weight
value of a load on the weighing platform.
By entering a percentage:
The upper value of Class 1 is entered
using the keypad or by saving the
value indicated. Upper limits for the
other classes are defined by entering a
percentage of deviation from the upper
limit of Class 1, using the keypad.
Example: Enter 100 g as the upper
limit of Class 1. Then enter 15%.
When working with 3 classes, this yields
the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g - maximum load
When working with 5 classes, this yields
the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g – 130 g
Class 4: >130 g – 145 g
Class 5: >145 g - maximum load
The initialization data remains valid
until deleted by pressing the c key
or until overwritten by a new value.
They remain saved after the scale is
switched off.
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until appl is
displayed
§ Confirm appl: Press the ) key
§ Select the Classification application:
Press the k key repeatedly and
confirm with the ) key
w
Classification Application Parameters
3.6. Minimum Load for Initialization
and Defining the Class 1 Lower
Limit
3.6.1* 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
4.3. Control output for SET as:
4.3.1* SET output
4.3.2 Ready to operate for
process control systems
4.7. Activation of outputs
4.7.1 Off
4.7.2 Always on
4.7.3* On at stability
4.8. Number of classes
4.8.1* 3 classes
4.8.2 5 classes
4.9. Parameter input
4.9.1* Weight values
4.9.2 Percentage
4.10. Automatic printing
4.10.1* Off
4.10.2 On
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times.