Example:
Determining the number of uncounted parts.
Settings (changes in the factory settings required for this example):
Application parameters: Application: Counting
Setup: Printout:
PrtProt: 7.6.; then select menu line item of your choice
(see “Configuration” for options)
1 Place empty container
on the platform
)
2 Tare the scale
Note: If the automatic tare
function is enabled, you do
not need to press the )
key to tare the scale; the tare
weight is saved automatically
when you place the container
on the platform.
3 Place a number of parts in
the container for the reference
quantity (in this example,
20 pcs)
20
4 Enter the number of parts
using the keypad
r
5 Activate calculation of refer-
ence sample weight
If the weight is too light, Reduce the minimum load set-
Inf 29 is shown in the main ting or increase the reference
display. sample quantity and the num-
ber of parts in the container.
6 Add more parts to the
container
Read the result
Opt is displayed if automatic
reference sample updating
is enabled
p
7 Print the results
nRef + 38 pcs Configured printout:
wRef + 0.003280 kg see page 80
G# + 0.373 kg
T + 0.248 kg
N + 0.125 kg
Qnt 38 pcs
--------------------