Example:
Measuring 25 m electrical cable.
Settings (changes in the factory settings required for this example):
Application parameters: Application: Neutral Measurement
Setup: Printout:
PrtProt: 7.6.; then select menu line item of your choice
(see “Configuration” for options)
1 Place empty container on the
platform
)
2 Tare the scale
Note: If the automatic tare
function is enabled, you do
not need to press the ) key
to tare the scale; the tare
weight is saved automatically
when you place the container
on the platform
Signum 3 only:
248
3 Enter the weight of 1 meter
of cable using the keypad
(in this example, 248 g)
O
4 Save value entered as
reference weight.
5 Place the desired amount of
cable in the container
Read the result
p
6 Print the result
nRef + 1 o Configured printout:
wRef + 0.248 kg see page 80
G# + 6.794 kg
T + 0.541 kg
N + 6.253 kg
Qnt 25 o
--------------------