Example:
Defining three classes.
Settings (changes in the factory settings required for this example):
Applications: application 2: Classification
Setup: Printout:
PrtProt: 7.6.; then select the menu line item of your choice
(see “Configuration” for options)
O
1 Begin input of class delimiters
Signum 3 only:
110
2 Enter the upper limit for
Class 1 via the keypad
(in this example, 0.110 g)
O
3 Save the upper limit for
Class 1
Signum 3 only:
130
4 Enter the upper limit for
Class 2 via the keypad
(in this example, 0.130 g)
O
5 Save the upper limit for
Class 2
6 Place the sample on the
weighing platform
Read the result
p
7 Print the result
Note: If automatic printout
of results is enabled, you do
not need to press the p key;
the results are printed auto-
matically.
Lim1 + 0.110 kg Configured printout:
Lim2 + 0.130 kg see page 80
G# + 0.118 kg
T + 0.000 kg
N + 0.118 g
class 2
-------------------
Application 2: Classification W