03015/03017
33
Continuity test:
Range Introduction
When the resistance of the measured circuit is less than about 20Ω, the built-in buzzer will sound.
When the resistance of the measured circuit is greater than 150Ω, the built-in buzzer will keep mute.
If the resistance of the measured circuit is between 20Ω and 150Ω, the buzzer may sound or keep mute.
4.000μF 1nF ±(3.5%+5
40.00μF 10nF
±(4.0%+5
400.0μF 100nF ±(5.0%+5
1000μF 1μF 仅供参考
Temperature only applies to 03017:
Temperature sensor: Type K thermocouple
Range Resolution Accuracy
-20℃~ 0
0.1℃
±(6.0%+5℃)
0℃~ 400 0.1 ±(1.5%+4℃)
400℃~ 1000 1 ±(1.8%+5℃)
-4 ℉~ 32 0.1 ±(6.0%+9 ℉)
32 ℉~ 752 0.1 ±(1.5%+7.2 ℉)
752 ℉~ 1832 1 ±(1.8%+9 ℉)
Note:
1 The accuracy specified in the table excludes the error of the thermocouple.
2The accuracy in the table is obtained under the assumption that the change of the ambient temperature is within
±1 . In the case of ambient temperature change of 5 , the rated accuracy will apply 1 h later only.
3With the help of the internal sensing device of the meter, the accuracy is guaranteed when the operating temperature
is between 18-28 . When the operating temperature of the meter is outside this range the accuracy cannot be
guaranteed.
4Make sure the fuse is intact before the thermocouple is used in the measurement.