Notes on logging
Logging ends if the temperature measurement fails. In this state, “Er”
appears on the display in the temperature section of the indicator LCD.
The condition that caused the indicator to stop logging is “Abnormal pulses
detected”.
Note that the SK-L200TH II
α
ends logging only if the probe is removed
from the indicator.
Notes on data downloading
Download the stored data according to the logging option selected.
(1) For the “Once” (default) and the “Ends at...” (limited amount) options:
From the Transfer (T) menu, click
Download
(
D
)
.
(2) For the Repeatedly (overwriting) option:
Download in the same way as for (1) above, and enter the date and
time logging ended.
Note: Take a note of the date and time at which the logging ended.
(3) For the “By pages” (partitioning) option:
From the Transfer (T) menu, click
Readout Only
(
R
)
to display informa-
tion on the data stored on the indicator. Click the page number to be
downloaded, and then click
Download
(
D
)
.
Note: Select one page at a time; you cannot download all pages at
once.