IPM25 User Manual
20
You will be prompted with InPut LO, press
Now enter the zero offset sensitivity of the sensor by using the
and button to select a digit and the and
buttons to change the digit. When the value has been set press .If you do not know this, simply enter all
zeros.
You will be prompted with dISP LO. Press to enter the display value required for the low input figure entered. The
value can be entered by using the
and button to select a digit and the and buttons to change the
digit. When the value has been set press
You will be prompted with InPut HI, press
Now enter the sensitivity figure supplied for the sensor by using the
and button to select a digit and the
and
buttons to change the digit. When the value has been set press .
You then be prompted with dISP HI. Press to enter the display value required for the high input figure entered.
The value can be entered by using the
and button to select a digit and the and buttons to change
the digit. When the value has been set press
You should now see donE displayed. This means the calibration was successful, press to the IPM25 to normal
operation mode, with the new calibration data stored. If you see FaiLEd, then you will need to repeat the calibration,
checking that you have completed the procedure in the correct order, and that the sensor is connected correctly.