EasyManua.ls Logo

Scaime IPM25 - Millivolt Per Volt Calibration Menu

Scaime IPM25
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IPM25 User Manual
14
Millivolt per Volt Calibration Menu
To enter the MilliVolt per Volt Calibration Menu, press and hold
and
buttons for 10 seconds
Parameter Set-up Information
5.0 gAIn
Press
To skip to next menu item
Press
To change the 5mV/V gain.
Here the factory gain calibration can be changed to a measured value (see Milli-Volt
Calibration Procedure – back end of manual).
Once the derived value has been entered Press
to confirm.
5.0 OFFS
Press
to skip to next menu item
Press
to change the 5mV/V offset.
Here the factory offset value can be changed to a measured value (see Milli-Volt Calibration
Procedure – back end of manual).
Once the derived value has been entered Press
to confirm.
50 gAIn
THIS CAN ONLY BE SET WHEN USING 50mV/V RANGE
Press
To skip to next menu item
Press
To change the 50mV/V gain.
Here the factory gain calibration can be changed to a measured value (see Milli-Volt
Calibration Procedure – back end of manual).
Once the derived value has been entered Press
to confirm.
50 OFFS
THIS CAN ONLY BE SET WHEN USING 50mV/V RANGE
Press
to skip to next menu item
Press
to change the 5mV/V offset.
Here the factory offset value can be changed to a measured value (see Milli-Volt Calibration
Procedure – back end of manual).
Once the derived value has been entered Press
to confirm.

Related product manuals