6.0 Frequently Asked Questions |17
6.0 Frequently Asked Questions |17
6.0 Frequently Asked Questions
6.1 No movement
Check Scientifica logo is illuminated on the front of the rack
• Check for correct switch position on side of cube for controlling
manipulator
• Check all cables and connectors for damage or loose connections
6.2 Drift
Allow the system to warm up
With most electrical components on the rig needing a warm-up
period, it is good practice to allow a certain warming-up time
after switching on the power of the set-up (up to one hour).
Check all bolts are fully tightened
Is the drift detected with the perfusion running?
(test for longer than 10 minutes)
Perfusion solution with a temperature different to the room
can lead to thermal expansion of the pipette – leave some
time for the pipette to adjust to new temperatures.
Fast perfusion may cause drift of the pipette.
Ensure the vacuum source is strong enough to break the
surface tension of the fluid in the bath
Is the microscope firmly mounted to the table top?
Many microscopes have rubber feet underneath. These may
degenerate with time and cause the microscope to be
unstable. Ideally the microscope should be bolted down onto
the table top.
Is the cooling headstage switched off?
Some amplifiers have a cooling facility on the headstage
(which is normally located on the amplifier front or back