12
WARNING: Do not tamper with any fixings other than those indicated in the
manual for mounting purposes.
There are no user serviceable parts within the UMS/XY Stage
Do not use an inadequately rated mains power cable for the UMS/XY Stage
Control Rack other than those specified in this manual.
If the detachable mains cable is replaced care must be taken to ensure that a replacement
cable of adequate rating is used to ensure damage to the user or system does not occur.
Replacement cable provided must meet local electrical regulations and be sufficient for:
100-240volts
50-60Hz
5 amps
Fused- T1.6A L 250v
5. 1 Cleaning
WARNING: Switch off and disconnect the mains supply before cleaning the
unit. There is no integrated mains disconnect located within the UMS/XY Stage.
Mains power must be disconnected by switching off the mains power supply
outlet and disconnecting the mains power supply cable.
Compressed air can be used to remove dust from the top of the UMS/XY Stage. Care must
be taken to prevent dust entering the unit during the process.
Never use fluid to clean the unit or allow fluid to enter the unit to prevent danger of
electrocution.
Do not use any type of abrasive pad or material to clean the unit.
6.0 Frequently Asked Questions
6.1 No movement?
Check Scientifica Control rack is properly connected according to the Control rack
User Manual and that the rack is turned on.
Ensure all cables and devices are correctly and securely connected as per the
relevant Control Rack User Manual and the rack is turned on.
Ensure that the correct device is selected using the device selection toggle switch on
the side of the Control Cube/PatchPad Display.