SD GlucoNavii® NFC BLOOD GLUCOSE MONITORING SYSTEM
Maintenance and Troubleshooting
61
6. Warnings, Precautions and Limitations
t Never make signicant changes to your diabetes
control program or ignore physical symptom without
consulting with your healthcare professional.
t Severe dehydration (excessive water loss) may cause
false low results. If you believe you are suering from
dehydration, consult your healthcare professional
right away.
t Extremes in hematocrit may aect test results.
Hematocrit levels greater than 70% may cause falsely
low readings.
t Inaccurate results may occur in severely hypotensive
individuals or patients in shock. Inaccurate low
results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without
ketosis. Critically ill patients should not be tested with
blood glucose meters.
t Normal endogenous (within body) natural levels
of uric acid, ascorbic acid (vitamin C), bilirubin,
triglycerides, and hemoglobin do not interfere with
your blood glucose results obtained.
t Interferences : The following compounds, elevated
levels of ascorbic acid, uric acid, acetaminophen, total
bilirubin, triglycerides may aect results.
Compound Limitation
Ascorbic acid > 4 mg/dL
Uric acid
> 9 mg/dL
Acetaminophen
> 6 mg/dL
Total bilirubin
> 40 mg/dL
Triglycerides
> 1800 mg/dL