How to use STANDARD Ezi Tube+ *Only for lipid testing
1. Touch the tip of a STANDARD Ezi Tube+ to the edge of the blood
drop. Blood will automatically be drawn up the tube by capillary
action. Keep the tube in contact with the blood until it is  lled to
the black line. You should squeeze your  nger if you need more
blood.
2. Transfer the collected sample to the sample well on the test strip
by gently squeezing the bulb of the STANDARD Ezi Tube+.
3. Close the  ap to start the test.
Used STANDARD Ezi Tube+ can cause infection. Please discard the used STANDARD Ezi Tube+ according to local regulations.
Collect the blood sample carefully to try and avoid air bubbles.
How to use the micropipette *Only for lipid testing
1 2
1
2
3
2
3
4
1
2
3
2
3
4
2
3
4
3
Place the tip securely on the micropipette. Push the plunger down and hold it in this position. Touch the tip to the blood drop.
4
5 6
Slowly release the pluger and check the blood
is drawn into the tip.
Open the  ap and apply the blood sample.
Remove the tip from the micropipette and dispose
of the tip in a biohazardous waste container.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SD Biosensor STANDARD LipidoCare and is the answer not in the manual?

SD Biosensor STANDARD LipidoCare Specifications

General IconGeneral
BrandSD Biosensor
ModelSTANDARD LipidoCare
CategoryMedical Equipment
LanguageEnglish