2.1.6 Example: Adjusting a Scale with 100g Graduations
[703|704|284|285|634|635|644|645|656|657|664|665|675|676|677
|684|685]
The selected test weight is 200 kg.
The current weight shown on the display is 199.7 kg (max. deviation of 300 g).
In this example, you must switch the scale to incrementing mode to carry out
readjustment. When the tare/hold key has been pressed (for more than 1.5
seconds), "inc" starts flashing.
Now increase the value by the first 10 g (by briefly pressing the tare/hold key).
After this has been repeated x times, the value 200.0 (rounded!) appears on the
display although the internal measured value still is 199.95 kg (not rounded!).
To correct this rounding error, increment the value five times by 10 g to obtain
exactly 200.00 as the calculated value.
Complete the readjustment procedure as described in section 2.1.5.
2.1.7 Example: Adjusting a Baby Scale with 10g Graduations
[374|376]
The selected test weight is 15 kg. The current weight shown on the display is
14.970 kg (max. deviation of 30 g).
In this example, you must switch the scale to incrementing mode to carry out
readjustment. When the tare/hold key has been pressed (for more than 1.5
seconds), "inc" starts flashing.
Now increase the value by the first 1 g (by briefly pressing the tare/hold key).
Each alteration is shown on the display.
Complete the readjustment procedure as described in section 2.1.5.