Errors seca
14.11.12/RLEU/ARI/MRE 21/40 30-34-00-812c
3 Errors
[374|376|378|703|704|634|635|644|645|656|657|664|665|
674|675|676|677|684|685|954|957|959|963]
The following section provides an overview of possible error symptoms, their
causes and the steps required to remedy the error. In addition, the error
messages generated by the device and ways to eliminate them are explained.
3.1 Error Symptoms for Scale
[374|376|378|703|704|634|635|644|645|656|657|664|665|674|675|
676|677|684|685|954|957|959|963]
Error description
Possible causes Remedy
Battery not inserted or flat
(if provided)
Check the batteries
Voltage supply defective See section 4.1
Scale does not
start
Key panel defective or not
connected
Check cable
connections, check
cable and key panel
for interruptions,
replace if necessary
Packaging / transport
locking device not removed
completely
Check the scale
Scale not correctly adjusted See section 2.1
Load cell damaged
See replacement
instructions
Scale shows no
weight or an
incorrect weight
Force transmission into
load cell interrupted
Check base frame for
damage or incorrect
assembly
Measured values
vary greatly
Measurement
electronics/load cell
damaged
See replacement
instructions