Seca 644 Series - 4.2.1 Measurement Using a Multimeter

Seca 644 Series
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements seca
14.11.12/RLEU/ARI/MRE 28/40 30-34-00-812c
Measurement between Ohmic resistance []
V
+
and Sig
+
4 (blue and red) ~250-300
V- and Sig
+
(blue and white) ~250-300
V
+
and Sig- (red and black) ~250-300
V
-
and Sig- (white and black)
~250-300
Sig
+
/Sig-/V
+
/V- and aluminium
body of the load cell
> 1000 M
Table 1: Resistance of platform load cell type Kogean / HBM [703|704x|954|957|959|963]
Measurement between Ohmic resistance []
V
+
and Sig
+
4 (green and red)
~ 800-850
V- and Sig
+
(green and white)
~ 800-850
V
+
and Sig- (red and black)
~ 800-850
V
-
and Sig- (white and black)
~ 800-850
Sig
+
/Sig-/V
+
/V- and aluminium
body of the load cell
> 500 M
Table 2: Resistance of Flintec / Kogean Planar Beam / --Wing load cell [376|378|284|285]
The relevant cabling diagram shows the designations and associated cable
colour. The measured values only have a limited informative value, however, as
even a defective load cell can supply absolutely correct values when no load is
placed on it. If the load cell supplies incorrect values when unloaded, it is
definitely defective.