Model 644 33
GB
6. Moving the scale
For mobile deployment in a ward, the scale
can be tipped backwards and moved
about on the rollers.
Important!
The alignment of the base of the scale
must be checked and corrected if nec-
essary every time the location of the
scale is changed.
7. Maintenance
Clean the rubber surface and the housing as required using a domestic cleaning agent or
commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for
cleaning. Such substances can damage the high-quality surfaces.
8. What do I do if…
… no weight display comes on?
Is the scale switched on?
Check the mains connection and the connection of the battery pack.
Is the plug contact for connecting the display to the scale correctly connected?
Switch the scale off for a few seconds. This can be done by opening the plug connect-
ed to the rechargeable battery block and unplugging the mains supply from the scale.
0.0 does not appear before the weighing operation?
Press the Start key again – there must not be any load on the scale – and only its feet
should be in contact with the floor.
… one segment is illuminated constantly or not at all?
The relevant segment has a fault. Inform Servicing.
… the display shows ?
Battery voltage is dropping. Charge the rechargeable batteries in the next few days.