80 •
… is displayed?
The ambient temperature of the scales is too high
or too low.
-Set up scales in an ambient temperature
between +10 °C and +40 °C
-Wait approx. 15 minutes until scales have
adjusted to ambient temperature
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
•The instrument was not able to send any mea-
surement results to the wireless receiver (seca
wireless printer or PC with seca USB wireless
adapter).
–Check that the scales are integrated in
the wireless network.
–Check that the receiver is switched on.
•Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
-Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.
... only the item "SYS" is
visible in the rf menu?
•The wireless module is disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 75).
... only the items "SYS"
and "lrn" are visible in
the rf menu?
•The wireless module is activated but no
wireless group set up.
-Set up wireless group (see "Set up wireless
group (Lrn)" on page 75).
... the items "APrt" and
"Time" are not visible in
the rf menu?
•No wireless printer is logged into the wireless
group.
-Use menu item "lrn" to log the wireless printer
into the wireless group (see "Set up wireless
group (Lrn)" on page 75).
…after calling up the
menu, the “rf” element is
not displayed.
•The scale’s wireless module is defective.
-Inform the maintenance service
…
Er:X:11 is displayed?
The scales are too high, or too great a load is
applied at one point.
-Reduce load on scales or distribute weight-
more evenly
-Restart scales
Malfunction Cause/Remedy