82 •
Height of figures 25 mm
Power supply
Rechargeable battery block
Power pack
Power consumption
•with wireless module disabled
•with wireless module activated
approx. 32 mA
approx. 50 mA
Maximum runtime in rechargeable battery
operation
•with wireless module disabled
•with wireless module activated
approx. 2.400 minutes
Power supply unit recom-
mended
Medical device according to directive
93/42/EEC
Class I with measuring function
EN 60601-1:
•Class II totally insulated appliance:
•Type B electromedical device:
Maximum load
•Partial weighing range1
•Partial weighing range2
150 kg / 330 lbs / 24 sts
300 kg / 660 lbs / 47 sts
Minimum load
•Partial weighing range1
•Partial weighing range2
1,0 kg
2,0 kg
Graduations
•Partial weighing range1
•Partial weighing range2
50 g / 0.1 lbs
100 g / 0.2 lbs
Tare range 300 kg
Accuracy
•0 to 35 kg
•35 kg to maximum load
•0 to 75 lbs
•75 lbs to maximum load
•0 to 5.5 sts
•5.5 sts to maximum load
±100 g
±0,3%
±0.2 lbs
±0,3%
±0.2 lbs
±0,3%
Radio transmission
•Frequency band
•Transmission power
•Standards applied
2,433 GHz - 2,480 GHz
<10 mW
EN 300 328
EN 301489-1
EN 301489-17, -177
Technical data for
seca 954