Connecting the power supply unit The rechargeable battery compartment is charged with the included power
supply unit.
WARNING!
Personal injury and damage to device if incorrect power supply unit is
used
The voltage provided by standard power supply units may be higher
than their indicated rating. This may cause the scales to overheat,
catch fire, melt or short-circuit.
Only use genuine seca plug-in power supply units with 9 V or a
controlled 12 V output voltage.
1. Insert the power plug necessary for your power supply in the power
supply unit.
2. Insert the connector plug of the power supply unit in the connecting
socket of the scales.
3. Plug the power supply unit into a mains socket.
4. Allow the scale to charge on the network for at least 24 hours when it is
first charged so that the rechargeable battery block is completely
charged.
4.3 Aligning the scale
1. Place the scale on firm and even ground.
2. Align the scale:
with the foot screws on the front transport rollers
with the knurled head screws on the rear transport rollers
The air bubble in the spirit level must be located in the centre of the circle.
Air bubble
Spirit level