Hygiene treatment • 21
English
17-10-07-602-002d_10-2019S
NOTE
If 0” is set, then if you are weighing patients who are not very steady
on their feet, it is possible that despite the “Hold” function being acti-
vated, no weight value will be permanently displayed.
The greatest deviation between the weight value displayed and the
actual value occurs with setting “2”.
3. Confirm the selection.
You will leave the menu automatically.
Restoring factory settings (rESEt) You can restore the factory settings for the following functions:
1. Select the item “rESEt” in the menu.
2. Confirm the selection.
You will leave the menu automatically.
3. Switch the scales off.
The factory settings are restored and will be available when the scales are
next switched on.
6. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
Ensure that the device is switched off before performing any hygiene
treatment.
Disconnect the power supply connector before performing any
hygiene treatment.
Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
Do not use caustic or abrasive detergents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).
6.1 Cleaning
Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
Function Factory setting
Autohold (AHOLd) Off
Acoustic signal (PrESS Off
Acoustic signal (HOLD) On
Attenuation (FIL) 0
Autoclear (ACLr) On
Pre-Tare (Pt) 0 kg
Height for Body Mass Index
(BMI)
170 cm