2.2 Basic safety precautions
Handling the device Please take note of the information in these instructions for use.
Keep the instructions for use in a safe place. The instructions for use are a
component of the device and must be available at all times.
DANGER!
Risk of explosion
Do not use the device in an environment in which one of the following
gases has accumulated:
oxygen
flammable anesthetics
other flammable substances/air mixtures
CAUTION!
Patient hazard, damage to device
Additional devices which are connected to electrical medical devic-
es must provide evidence of compliance with the relevant IEC or ISO
standards (e.g. IEC 60950 for data-processing devices). Further-
more, all configurations must comply with the requirements of stan-
dards for medical systems (see IEC 60601-1-1 or Section 16 of
edition 3.1 of IEC 60601-1, respectively). Anyone connecting addi-
tional devices to electrical medical devices is considered a system
configurer and is therefore responsible for ensuring that the system
complies with the requirements of standards for systems. Your at-
tention is drawn to the fact that local laws take precedence over the
above-mentioned requirements of standards. In the event of any
queries, please contact your local specialist dealer or Technical Ser-
vice.
Have servicing and verification carried out regularly as described in
the relevant section of this document.
Technical modifications may not be made to the device. The device
does not contain any parts for servicing by the user. Only have ser-
vicing and repairs performed by an authorized seca service partner.
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
Only use original seca accessories and spare parts, otherwise seca
will not grant any warranty.
CAUTION!
Patient hazard, malfunction
Keep other electrical medical devices, e.g. high-frequency surgical
devices, a minimum distance of approx. 1 meter away to prevent in-
correct measurements or wireless transmission interference.
Keep HF devices such as cell phones a minimum distance of ap-
prox. 1 meter away to prevent incorrect measurements or wireless
transmission interference.
The actual transmission output of HF equipment may require mini-
mum distances of more than 1 meter. Details can be found at
www.seca.com.
CAUTION!
Patient hazard, damage to device
Use the chair scale only under the supervision of specialist staff.
Prior to each use, ensure that the wheel brakes work properly.
Secure the scales against accidental rolling. Lock the wheel brakes.
Do not transport any persons or objects with the chair scales.
Move the scales only with the wheel brakes released.