EasyManua.ls Logo

Seintek S2401 - Entering Calibration Mode; Calibration Scope Center

Seintek S2401
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-7
4.3. Entering Calibration Mode
With retaining ,
buttons pressed, press
button.
Examine the screen status
and buttons function.
Key: , , + Simultaneously.
4.4. Calibration Scope Center
1. Connect the tool for a short circuit test to CHA, CHB, and COM
connector of the equipment.
2. Adjust Scope center for each V/DIV. You can change V/DIV by
pressing up-down arrow buttons of the equipment. (V/DIV of CHA,
CHB will be changed simultaneously when you adjust them.)
3. Move to 0.5V/DIV.
4. Scope center would be adjusted by pressing F3 button (Z/cal.). If
the equipment still do not display scope center correctly, press F3
(Z/cal.) continuously until the equipment displays the correct scope
center.
5. Adjust the scope center of 50mV, 0.1V, 0.2V, 0.5V, 1V, 2V, 5V, 10V,
20V, 50V, 100V, 200V, 500V as described above. (50mV, 0.1V,
0.2V is only S2405, S2505)

Table of Contents

Related product manuals