11 Protective measures and precautions
42
The manufacturer does not assume any liability for damages
caused by improper or unintended use.
11.3 Hazard areas and residual hazards
The sensors are installed in and/or on tanks and pipelines that
may be pressurised. Process liquid may only escape in the event
of negligent conduct and improper operation.
Before commissioning and after each maintenance activity,
ensure that all gaskets and connections are complete and
functional.
Take appropriate protective measures before making contact
with the sensor, because the parts may adopt the process
temperature.
11.4 Equipment
Only use tested and approved accessories and equipment.
The STS 03 sensor does not require any elastomer gaskets on the
process connections. In case you connect the sensor to your
process by means of an adapter,
select the material properties of the process gaskets and the
O-rings subject to the process medium and the flushing fluid.
take into consideration the sealing material’s capacity to
swell and its resistance against acids and/or alkaline
solutions.
11.5 Personnel
Only trained technical personnel must install and maintain the
sensor!
!!!