18 Troubleshooting
18.1 No or erroneous measured value
possible cause
remedy
No voltage at the sensor
Check/make electrical connection in
accordance with (chapter 4.3)
Measuring windows are coated
Clean the measuring windows
(chapter 7.3)
If the measuring windows get dirty
frequently, it is better to use an
STS04 with SAW8XX quick-change
fitting.
Erroneous calibration
Reset the calibration to factory
settings in the calibration menu
(chapter 6.1)
18.2 Strongly fluctuating measured value
possible cause
remedy
Air bubbles in the system
Dampen display and output current
(chapter 5.2)
Sensor is not completely submerged
in the process fluid
Change the installation location