18 Troubleshooting
72
18.3 Output current does not match the measured value
possible cause
remedy
Current output parameterised
incorrectly
Check and, if required, change the
parameterisation of the current
output (chapter 5.2)
Electrical connection incorrect
Check/make electrical connection in
accordance with (chapter 4.3)
18.4 Switching output does not switch properly
possible cause
remedy
Switching output parameterised
incorrectly
Check and, if required, change the
parameterisation of the switching
output(chapter 5.3)
Electrical connection incorrect
Check/make electrical connection in
accordance with (chapter 4.3)
18.5 Keyboard cannot be operated
possible cause
remedy
Keylock is activated
Unlock the keyboard (chapter 5.5)