18 Troubleshooting
73
18.6 Error message on the display
Err0
Factory comparison failed
Repair with the manufacturer
Err1
User parameters incorrect
Use the RST function in order to
reset the user parameters (chapter
5.7)
Err2
Temperature measurement failed
Repair with the manufacturer
Err3
Measuring range spreading failed
Check the parameters “MRB” and
“MRE”. Maximum spreading 4:1
Repair with the manufacturer
Err4
Temperature overriding (100°C)
Factory setting
Lower the temperature of the medium
below 100 ° C again