12 Product description
47
(UnIt) Defines the displayed unit of the measured value. For the
absorption measurement, it is possible to select between AU
(absorption unit) and a customer-defined, free measuring unit
CDU (Customer Defined Unit).
If you selected the opacity measurement option as measuring
principle, you can select between the following measuring units:
EBC European Brewery Convention
FAU Formazine Absorption Unit
TEF Trübungseinheiten Formazin (opacity units
formazine)
MGL Milligrams per Litre
and a customer-defined, free measuring unit CDU (Customer
Defined Unit).
In this, the following is applicable: 1 FAU = 1 TEF = 0.25 EBC =
2.05mg/L.
(CdUd) Defines the decimal point (decimal place) in the display
(CdU) Defines the display scope of the user unit
Defines which measured value is to be displayed:
(turB) Opacity / absorption
(tEMP) Temperature
(Alt) Opacity / absorption and temperature alternating
Regardless of the display switchover, the analogue output always
delivers a signal depending on the opacity / absorption.
(Mrb) Defines the 4mA point for the output current. The range can
be selected freely between 0 …19999 (0.000…19.999). The
measuring range is set in the measuring unit currently used in
each case.
(MrE) Defines the 20mA point for the output current. The range
can be selected freely between 0 …19999 (0.000…19.999). The
measuring range is set in the measuring unit currently used in
each case.
(dAN) Damps the measured value in the range of 0.0…200.0
seconds both for the output current and for the display.
(r-0-) Defines a range in display digits around the zero point where
the measured value is set to 0.
Measuring unit