EasyManua.ls Logo

Sensorex TX100 - TX100 Specifications

Sensorex TX100
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRODUCT INSTRUCTION MANUAL
Page 2 of 35
1.3 TX100 Specications
Measuring Range (pH) 0.00 to 14.00pH, 0.01 pH resolution,
+/- 0.01 accuracy
Measuring Range (ORP/mV) -1999mV to 1999mV, 1mV resolution,
+/- 2mV accuracy
Measuring Temperature Range -20 degC to 110 degC /-4 degF to 230degF,
0.1degC/F resolution
Current Output Range 2.00mA to 24mA (4-20mA galvanically isolated) , 0.01mA
resolution,+/-0.005mA accuracy
Enclosure NEMA 4X, IP65, ABS case with silicone keypad
HWD: 3.8” (96.52 mm) x 3.8” (96.52 mm) x 2.8” (71.5mm)
Weight approx. 1lb (.45kg)
Mounting Options Wall mount, panel mount, pipe mount and DIN rail (rail not
included)
Conduit Openings Standard: 2 - 3/8” openings cordgrips included.
Ambient Temperature Transmitter Service, 0 degC to 60 degC / 32 degF to 140 degF
Sensor Service – Refer to Sensor specications
Ambient Humidity 0 to 95% (non-condensing)
Location Designed for non-hazardous areas
Temperature Input 2-wire Pt1000 RTD with automatic compensation
Max. Sensor-to-Transmitter Distance 30 feet (9.1 meters)
Power 12-24 V .DC, 8 amp maximum current
EMI/RF EN 61326-1
Supply Voltage ( V DC) Max Resistance Load (Ohms)
12 150
16 350
20 550
24 750

Related product manuals