32
Setra Super Count
TM
Operator's Manual
PROGRAMMING A MINIMUM SAMPLE SIZE
This feature allows you to preset the minimum number of pieces that will be requested when the
sample size key is pressed. To enter the minimum sample size you wish to have used when counting,
use the following procedure:
1. Press the — key.
2. Enter 2222.
3. Press the setup/cal key. The display will read “PC nnn”, where “nnn” is the current minimum
sample size setting.
4. Enter the minimum number of pieces you wish to be used.
5. Press the setup/cal key.
NOTE: To discontinue use of the Minimum-Sample-Size feature, repeat this procedure and enter 0 in
Step #4.
PROGRAMMING A MINIMUM ACCURACY
This feature allows you to preset the minimum accuracy level used by the scale based on the sample
weight. To enter a specific minimum accuracy level, use the following procedure:
1. Press the — key.
2. Enter 3333.
3. Press the setup/cal key. The display will read “ACnn.nn”, where “nn.nn” is the current mini-
mum accuracy setting.
4. Enter the desired accuracy in percent. (A number between 95 and 99.99, e.g., 99.9.)
5. Press the setup/cal key.
NOTE: For any given sample, the scale automatically calculates the number of pieces needed to be
added to the sample, if any, based upon the sample weight necessary to achieve a desired
accuracy. The higher the accuracy selected, the larger the sample weight the scale will re-
quest. To discontinue use of the Minimum Accuracy feature, repeat this procedure and enter
0 at Step #4.
Programming the minimum accuracy does not compensate for piece weight variations in the
sample. It is always recommended that a large enough sample is used when determining an
average piece weight.