PIECE WEIGHT PER THOUSAND
When counting extremely small parts, accuracy and efficiency can be increased if the APW is repre-
sentative of 1000 pieces rather than of a single piece. The piece weight per thousand mode allows
the scale to count tiny parts using an APW that is 1000 times the weight of one part. The factory
default setting for this mode is an average piece weight for a single piece. This mode is enabled
using the following steps:
1. Enter — 14443 and press the setup/cal key.
2. The scale will display “APW 1”.
3. To select an APW representative of a single piece press setup/cal.
To select the piece weight per thousand mode, press the clear key.
4. The scale will display “APW 1000”.
5. To select an APW representative of 1000 pieces, press the setup/cal key.
Press clear to exit this procedure without changing the setting of this mode.
CAUTION: Any APWs that were previously stored will be incorrect if this mode is changed. All APWs
will become representative of either one piece or 1000 pieces depending on how the
mode was set previous to the change.
APW/TARE SCALE DATABASE SETUP
This setup must be enabled before any ID’s and their associated APW and tare values may be stored.
The factory default for this setup is disabled.
NOTE: An ID’s associated APW may be stored with no tare value but a tare value may not be stored
without an APW.
To enable the scale database, use the following procedure:
1. Enter — 15554 and press the setup/cal key.
2. The scale will display “AUTO ? APW.”
3. Press the setup/cal key to store the APW value associated with the ID or press the clear key to
exit this setup.
4. The scale will display “AUTO ? TARE.”
5. Press the setup/cal key to store the tare value associated with the ID or if you do not wish to
store the tare value press the clear key.
6. The scale will now display “CLEAR.” The database is now enabled and can be used to store
ID’s and their associated APW and/or tare values.
If this is not the first time you have enabled the database feature then you may already have
values stored. Press the clear key if you do not want to lose any of these existing APW and/or
tare values.
Press the setup/cal key to clear all IDs and their APW and/or tare values that may have been
previously stored.