17
• this measuring range is used to
monitor proximity to the
LELLEL
LELLEL
LEL
(
ll
ll
lower
ee
ee
explosion
ll
ll
limit)
• after approx. 8 seconds' warm-
up time a stable zero point is
reached
• the pump runs at minimum power
and the acoustic operating signal
(item 7) sounds every 5 seconds
• when the pre-set threshold of
20 %LEL is exceeded
methane:methane:
methane:methane:
methane: 20 %LEL = 0,88 %VOL
(display = 0,90 %VOL)
propane:propane:
propane:propane:
propane: 20 %LEL = 0,34 %VOL
the
advance alarmadvance alarm
advance alarmadvance alarm
advance alarm is triggered
both optically (item 1) and
acoustically (item 7)
• the advance alarm is an
intermittent tone quite distinct
from the operating signal;
it
ceases after a certain periodceases after a certain period
ceases after a certain periodceases after a certain period
ceases after a certain period
and cannot be cleared
• if the concentration exceeds
100 %LEL
methane:methane:
methane:methane:
methane:100 %LEL = 4,40 %VOL
propane:propane:
propane:propane:
propane:100 %LEL = 1,70 %VOL
the
main alarmmain alarm
main alarmmain alarm
main alarm is triggered
optically (item 1) and acoustically
(item 7)