Note:
The device can be stored temporarily up to a temperature of
50°C, however prolonged exposure to temperatures above 45 °C
will reduce the battery's performance and life.
NOTICE! Danger of battery destruction
Rechargeable batteries which have completely lost their charge
cannot be recharged.
● Before storage, charge or discharge the rechargeable battery
to around 30 – 50 % capacity.
● Always store the battery in optimum storage conditions.
● Recharge the battery every 6 months in order to ensure it nev-
er self-discharges entirely. Only ever charge the rechargeable
battery to around 30 – 50 % capacity.
3.6.2.3 Charging the battery
The rechargeable battery is integrated into battery compartment
cover. There is a charging socket on the long side of the battery
compartment cover.
NOTICE! Danger of battery destruction
The rechargeable battery has a dedicated adapter.
● Only use the UT 9000 AC/DC adapter to charge the battery.
● Before using the AC/DC adapter to charge the battery, ensure
it is in perfect condition. Never use a damaged AC/DC adapter.
● Always observe the permitted charging conditions: tempera-
ture 0 °C – 45 °C, relative humidity < 80 %.
The battery compartment cover is locked with quick-release fas-
teners.
1.
Turn the quick-release fasteners to open the battery com-
partment. Remove the battery compartment cover with the
integrated rechargeable battery.