5-6-7 Analog Output (4~20mA, 0~10V) Check Mode
Analog output simulation : 0(4mV,0V)~20(20mV,10V)
If the range is 4~20mA, A will be displayed.
If the range is 0~10V, V will be displayed.
Use numeric and enter key to check the output value with unit 0.1.
If entered value is over maximum value, actual output will be 100% of that range.
EX) When analog output mode is 4~20mA, 4mA will be out if you enter 4.0
When analog output mode is 4~20mA, 20mA will be out if you enter 20.0
When analog output mode is 0~10V, 4.7V will be out if you enter 4.7
When analog output mode is 0~10V, 10V will be out if you enter 10.0