- 73 -
7-2. Calibration Process
Display
Cause
Treatment

When Max capacity/digit value is over
20,000
Re-input the Max Capacity, less than 20.00
(Max Capacity / Digit)

Standard weight value is over than Max
Capacity
Re-input Standard weight value with
Number keys, under Max Capacity

Standard weight value is less than 10% of
Max Capacity
Re-input Standard weight value with
Number keys, more than 10% of Max
Capacity

Amp. Gain is too big
Check standard weight’s weight with set
value.
If there is difference between set value and
real weight, please re-input the value
(set value is too small)

Amp. Gain is too small
Check standard weight’s weight with set
value.
If there is difference between set value and
real weight, please re-input the value
(set value is too big)

Under “F-function” model, set value is “N.A”
Check the correct value and re-input

Ethernet option function set error
Re-set Ethernet option card function. (F-function
405~417)

When there is continuous vibration on the
weighing part,, indicator cannot process
calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition
check
 /  are displayed under situation when it is hard to calculate span value by the