- 73 -
7-2. Calibration Process
When Max capacity/digit value is over
20,000
Re-input the Max Capacity, less than 20.00
(Max Capacity / Digit)
Standard weight value is over than Max
Capacity
Re-input Standard weight value with
Number keys, under Max Capacity
Standard weight value is less than 10% of
Max Capacity
Re-input Standard weight value with
Number keys, more than 10% of Max
Capacity
Check standard weight’s weight with set
value.
If there is difference between set value and
real weight, please re-input the value
(set value is too small)
Check standard weight’s weight with set
value.
If there is difference between set value and
real weight, please re-input the value
(set value is too big)
Under “F-function” model, set value is “N.A”
Check the correct value and re-input
Ethernet option function set error
Re-set Ethernet option card function. (F-function
405~417)
When there is continuous vibration on the
weighing part,, indicator cannot process
calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition
check
※ “ / are displayed under situation when it is hard to calculate span value by the
load.