6-1-7. Read COMMAND for SI410 (Under F-function 202/212-01)
Current Weight
ASCII : STX ID(2Byte) RCWT ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RCWT + ETX [8Byte]
Transmission: STX + ID(2Byte) + RCWT + State 1(1Byte) + State
2(1Byte) + P + Decimal Point(1Byte) + Mark(1Byte) +
Current weight(6Byte) + Unit(2Byte) +ETX [21Byte]
8 byte
21 byte
State1: O(Overload), S(Stable), U(Unstable)
State2: N(Net Weight), G(Gross Weight)
Current data
ASCII : STX ID(2Byte) RCWD ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RCWD + ETX [8Byte]
Transmission: STX + ID(2Byte) + RCWD + P + Decimal Point
(1Byte) + Date(6Byte) + Time(6Byte) + Part Number(2Byte) +
Count(6Byte) + Mark(1Byte) + Tare(6Byte) + Mark(1Byte) +
Current weight(6Byte) + Unit(2Byte) +ETX [46Byte]
8 byte
46 byte
Sub-total data
ASCII : STX ID(2Byte) RSUB ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RSUB + ETX [8Byte]
Transmission: STX + ID(2Byte) + RSUB + P + Decimal Point(1Byte)
+ Part Number(2Byte) + Sub-total Count(6Byte) +
Sub-total weight(10Byte) + Unit(2Byte) + ETX [30Byte]
8 byte
30 byte
Grand-total data
ASCII : STX ID(2Byte) RGRD ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RGRD + ETX [8Byte]
Transmission: STX + ID(2Byte) + RGRD + P + Decimal(1Byte) +
Grand-total Count(6Byte) + Grand-total weight(10Byte) +
Unit(2Byte) + ETX [28Byte]
8 byte
28 byte
Sub-total count
ASCII : STX ID(2Byte) RSNO ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RSNO + ETX [8Byte]
Transmission: STX + ID(2Byte) + RSNO + Sub-total Count(6Byte) +
ETX [14Byte]
8 byte
14 byte
Current time
ASCII : STX ID(2Byte) RTIM ETX
Length
SI 410
Response
Reception: STX + ID(2Byte) + RTIM + ETX [8Byte]
Transmission: STX + ID(2Byte) + RTIM + Time(6Byte) +
ETX [14Byte]
8 byte
14 byte