73 | P a g e
USER MAINTENANCE (CONTINUED)
End of procedure
Temperature Calibration (Continued)
8. If the two readings are not the same, and the difference still falls outside
the acceptable range of your protocol, repeat steps 3 7 up to two more
times.
a. Three calibration attempts may be required to successfully calibrate
units that are more than ±2°C out of calibration.
Reference Device
9. If the temperature readings of the incubator temperature display and the
reference device still fall outside your protocol after three calibration
attempts, contact your distributor or Sheldon Technical Support for
assistance.