TOCV-CP_IMshort_Control-ver-2_Rev07F Page 10 of 22
(Software version ≥ 2.00) effective date: September 2007
Calculation Method
Point to point
Linear regression
The point to point curve fit draws a straight line between adjacent
data points and considers each line segment to be a separate
calibration line governed by its own equation.
A linear regression statistically determines the line that best fits the
pattern of all data points.
Zero Shift
Select this item to shift the calibration curve through origin.
Multiple Injections
If this item is selected, multiple injections are made from a single
aliquot of standard solution. This larger standard aliquot is drawn in
the syringe in one aspiration an is sufficient to accommodate all of
the repeat injections
File Name
Enter the name of the new calibration curve. Click on the browse
button to save the file in a directory other than default directory
Calibration Curve Wizard 4 Calibration Measurement Parameters
Parameter
Action/Description
Units
Select the concentration unit.
Concentration Range
Enter the lower and upper limits of the calibration curve
concentration range
No. Cal. Points
Enter the number of calibration standards
No. of Injections
Enter the minimum/maximum number of injections for the standards
No. of Washes
Enter the number of times the syringes is washed with sample
SD Max.
Enter the standard deviation limit. This limit is used to assess the
analysis result
CV Max
Enter the coefficient of variation limit. This limit is used to assess the
analysis result
Sparge Time
Enter the desire sparge time. This option is available only for NPOC
analysis.
Acid Addition
Enter the percentage of acid to be added to the standard solution
Calibration Curve Wizard 5 Calibration Points List
[please refer to page 12]
Parameter
Action/Description
No.
Displays the calibration point number. Note that the standards will be
analysed in the order shown
Auto Dilution
Displays the calculated automatic dilution factor
Injection Volume
Displays the injection volume for the calibration standards
Calibration points
Displays a table of measurement parameters for each calibration point
[Edit]
To edit the parameter for calibration point, highlight the point in the
table and click on [Edit] button
[ADD]
To add a calibration point, click on the appropriate number in the No
column to specify the insertion point, then click on [ADD] button
[Delete]
To delete a calibration point, highlight the point in the table and click
on [Delete] button
[Delete All]
To delete all calibrations, click on [Delete All] button