TOCV-CP_IMshort_Control-ver-2_Rev07F Page 11 of 22
(Software version ≥ 2.00) effective date: September 2007
Calibration Curve Wizard –6 Additional Settings
Select this option to use the default values for peak detection.
Enter the minimum time the instrument signal will continue to be
detected
Enter the maximum time the instrument signal will be detected
when no peak is found
If this item is selected, the calibration curve error judgement is
based on the correlation coefficient. If the calibration curve contains
less than 3 calibration points, error judgement is not possible and
this option will have no effect
Failure Action (1
st
time)
Continue
Stop
Repeat
Select the process to be conducted if an error occurs.
Records the error and continue the measurement
Records the error and stops analysis
Records the error and re-analyse the calibration curve
Failure Action (2
nd
time)
Select the process to be conducted if a second error occurs.
Records the error and continue the measurement
Records the error and stops analysis
Enter the minimum correlation coefficient value. If the calculated
correlation coefficient is less than the value entered, an error is
assessed.
Click Finish. The calibration curve is saved.