TOCV-CP_IMshort_Control-ver-2_Rev07F Page 12 of 22
(Software version ≥ 2.00) effective date: September 2007
Overview about several possibilities to create a Calibration Curve:
Calibration points are
distributed uniformly over
concentration range
Edit calibration curve
point manually
Calibration curve
according to
Calibration curve according to
USP/EP
Normal calibration curve, data
points are added
automatically, helps by
multiple point calibration
curve
Normal calibration
curve, data points are
added manually
Special calibration curve:
always10 points, first point
and last point are injected
10times all other points only
one time; Linearity and
homogeneity are calculated
Special calibration curve:
Always 2 points:
0 and 500ppb
(United States Pharmacopoeia or
European Pharmacopoeia
Units
Concentration range
No. of Cal points
No of Injections
No of Washes
SD Max
CV Max
Sparge time
Acid addition
Selectable default: ppm
Enter range: e.g 0-5
Enter No of Points: e.g. 3
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Selectable default: ppm
Disable
Disable
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Selectable default: ppm
Enter range: e.g. 10-100
Recommended: 10
Recommended: 1
Selectable (0-10)
Disable
Disable
Selectable (0-20min)
Selectable (0-20%)
Recommended: ppb
Fixed conc. range:0-500
Fixed No. of Cal. Points: 2
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Calib. Points list
Inj. Volume
Calibration Points
Recommended value
Example: 0ppm
5ppm
10ppm
Calculated automatic.
Empty calibration
points list. Manually
filling via [ADD]
Recommended value
List of 10 calibration points
in the range from 10 – 100
(as example)
Recommended value
List: 0ppb
500ppb