TOCV-CP_IMshort_Control-ver-2_Rev07F Page 16 of 22
(Software version ≥ 2.00) effective date: September 2007
- Right-click in the left cell in the row where multiple samples are to be inserted. Select
“Insert” – “Multiple Samples”
Sample Group Wizard 1 Sample Source
- enter the parameter source for the sample
Parameter
Action/Description
Method
- to create the sample group using measurement parameters of an
existing method file
- Enter the desired method or select the method using the Browse
button
Calibration Curve
- to create the sample group using the measurement parameters of an
existing calibration curve
- Enter the desired calibration curve or select the curve using the
Browse button
Sample Group Wizard 2 Sample Parameters
Parameter
Action/Description
No of Samples
Enter the number of samples in the group. (1-100.)
Start Vial
Enter the starting vial position for the sample group. This field is
disabled if ASI is not supported by the system
Sample Name
Sample ID
The default designations for these fields are obtained from the method
or calibration curve identified in page 1. Enter other name if desired.
Index Start
Select this option to start counting index. Each sample added receives
an increased counting index number.
Insert Cal. Curve /
Control Samples
Select this option to insert calibration curves and control samples alon
with the samples
Click Finish to insert the group of samples into the sample table
If “Insert Cal. Curve / Control Samples was selected, the Next button is displayed to enable
proceeding to the next step in the wizard
Sample Group Wizard 3 Calibration Curves
- to define the sequence of calibration
Parameter
Action/Description
At the beginning of
the sample group
Select this item to insert a calibration curve before the sample analysis
Always, after
“number “ of
samples
Select this item to inert a calibration curve after every fixed number of
samples. Enter that number in the box
Calibration Curves
Displays the information related to the added calibration curves. Up to
3 separate calibration curves may be added
[Add]
Use this button to add a calibration curve
[Delete]
Highlight a calibration curve listed in the window and click on the
Delete button to remove the curve from the sample group