PM2.5 Cyclone Particle Size
Selector For LD-5
OPERATIONMANUAL
CODE080000-251
Thank you for purchasing this product.
z This operation manual describes precautions that are important for preventing accidents as well
as the procedures used to handle the product.
z Read this operation manual thoroughly before use, and use the product correctly.
z After reading this operation manual, keep it handy for future reference.

Questions and Answers

  • H
    hwaltonAug 15, 2025
    What to do if Sibata LD-5 measurement value is abnormal (high)?
    • K
      Karen MoranAug 15, 2025
      If you observe a high, abnormal measurement value on your Sibata Measuring Instruments, the sensitivity correction might be off. Try closing the air sampling port and performing the SPAN measurement several times after you've done the BG measurement. This may resolve the issue.
  • J
    Joseph GardnerAug 18, 2025
    Why Sibata LD-5 shows nothing after switching the power ON?
    • V
      Victoria MillsAug 18, 2025
      If your Sibata Measuring Instruments shows no display after powering on, it could be due to a couple of reasons. First, the AC adapter might not be properly connected, so check that connection. Second, the battery level might be too low, so consider replacing the batteries.
  • R
    robertjonesAug 21, 2025
    What to do if Sibata LD-5 Measuring Instruments measurement value is abnormal (low) or there is no count?
    • G
      George DavisAug 21, 2025
      If you're getting an abnormally low measurement value or no count on your Sibata Measuring Instruments, several factors could be at play. Here are a few things to check: * Ensure the air sampling port cap is open before measuring. * If the BG measurement value or sensitivity correction is abnormal, perform the SPAN measurement after the BG measurement. * Check if the pump filter is clogged. If so, replace it and reset the operating time. * Clean the air sampling port with a brush. If these steps don't resolve the issue, the laser may be malfunctioning or the suction pump may have failed, requiring repairs or calibration.
  • C
    Casey WilsonAug 25, 2025
    Why Sibata LD-5 Measuring Instruments bar graph display is unstable even when measurements are not performed?
    • A
      alexanderrichardsonAug 25, 2025
      If the bar graph display on your Sibata Measuring Instruments is unstable even when you're not taking measurements, it's possible that dust inside the detector is being misinterpreted as a measurement value. Try closing the air sampling port and performing the BG measurement several times. If this doesn't fix the issue, the unit might need an overhaul.
  • M
    myoungAug 27, 2025
    What to do if Sibata LD-5 LCD is OFF?
    • J
      jason77Aug 28, 2025
      If the LCD on your Sibata Measuring Instruments is off, it's likely that the LCD sleep function has been activated. Simply press any switch to restore the display.
  • Y
    yjohnsonAug 31, 2025
    Why Sibata Measuring Instruments measurement value does not change?
    • E
      Edward StoneAug 31, 2025
      If the measurement value on your Sibata Measuring Instruments isn't changing, the scattering plate might be inserted. Make sure to turn the scattering plate knob to the [MEASURE] position and that the air sampling port is open.
  • S
    Sarah SaundersSep 3, 2025
    What to do if Sibata Measuring Instruments LCD screen does not change or the content displayed is strange?
    • J
      Jennifer SullivanSep 3, 2025
      If the LCD screen on your Sibata Measuring Instruments isn't changing or the display is strange, it might be a temporary LCD issue. You can try the following: 1. Press the [ESCAPE] switch. 2. Press the [UP] or [DOWN] switch. 3. Press the [ENTER] switch. After doing one of these, switch to your desired mode. You can also try pressing and holding the [ESCAPE] switch for at least two seconds to turn the LCD off, then press any switch again to restore the display. Alternatively, turn the power switch off and then back on. If none of these steps work, the unit may need repairs.
  • J
    Jennifer FrancisSep 7, 2025
    How to fix the issue when Sibata LD-5 current time has been reset?
    • E
      Edward BurgessSep 7, 2025
      If the current time on your Sibata Measuring Instruments keeps resetting, the backup battery is probably depleted. Replace the backup lithium battery to resolve this.
  • J
    Jeremy JohnsonSep 12, 2025
    What to do if Sibata LD-5 data cannot be recorded?
    • J
      Justin JosephSep 12, 2025
      If your Sibata Measuring Instruments is unable to record data, it's likely that the memory is full. To fix this, go to the log setting mode and delete some of the existing data.

Sibata LD-5 Specifications

General IconGeneral
BrandSibata
ModelLD-5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals