What to do if SICK CDB620 ID sensor cannot be accessed via SOPAS ET?
- BballardkristinAug 4, 2025
If the ID sensor cannot be accessed using the SOPAS ET configuration software after connecting the PC to the “AUX” male connector, ensure that switch S 4 is set to the “NO” position if no CMC600 is plugged in. Additionally, reload and integrate a suitable device description file in SOPAS ET via the device catalog, and use the “Device search” function to establish communication with the ID sensor.