8016441/2013-11-22•©SICKAG•Subjecttochangewithoutnotice 31
Commissioning
7 Commissioning
7.1 Aligning the ISD400 Pro
1. BringtheconnecteddevicesF1andF2ascloseaspossibletoone
another.
b Deviceisinoperatingmode.Anestablishedopticalconnectionwith
thedeviceoppositeisindicatedbythereceiveindicator.
2. Atcloserange(<1m),alignthedevicewithastraight,mechanicaltool
suchasalongspiritlevelifnecessary,untilthegreenreceiveindicator
lightsup.
3. MovedevicesF1andF2awayfromoneanotheruntiltheconnectionis
lostandthegreenreceiveindicator3isashingononeofthem.
4. Realignthedevices.TousetheopticalalignmentaidondeviceF1
(ISD400-7xx1),switchoffthelaser:Press
and atthesame
time.
5. Readjustwiththealignmentsightandopticalalignmentaidwithcross-
hairsuntilthegreenreceiveindicatorlightsupondeviceF1.Ifthere
isalowreceptionlevel,theopposingsendermustberealigned.In
operatingmode,thequalityofthealignmentofthesendercanberead
directlyfromthedisplayofthereceiver.
6. SwitchthelaseronagainondeviceF1(ISD400-7xx1):Press
.
7. Checkthealignmentusingthelevelindicatorandne-tuneifnecessa-
rytoadjust.