Operating Instructions PBS plus
OPERATING INSTRUCTIONS PBS plus8024946 | SICK AG
Subject to change without notice
53
6. FAULTS
5.11 Measuring range scaling (turn down)
In variants with analog signal, the measuring range beginning and measuring range end can be
adjusted within the permissible measuring range (do not exceed turn down of 5:1).
5.12 Description of IO-Link functionality
IO-Link is a point-to-point connection for communication of the PBS plus to an IO-Link master.
IO-Link specication: Version 1.1
A detailed description of the IO-Link function as well as the device description le (IODD) can be
found online on the detailed product page of the pressure switch at www.sick.com.
6. Faults
CAUTION!
Bodily injury, property and environmental damage
If faults cannot be eliminated by the measures listed here, put the pressure switch out of
operation immediately.
Ensure that there is no pressure or signal present and protect against accidental
commissioning. If measuring range scaling is set, pressure can be present despite an
analog value of 4 mA or 0 V.
Contact the manufacturer.
If the product has to be returned, follow the instruction in chapter 8.2 “Returns”.