ACUSON Freestyle User Manual 3-5
Setup
Setup Menu Options
Scan
Activates the real-time imaging screen. When you are in a Setup page or menu the SETUP Soft
Key changes to SCAN.
■ To go to the real-time imaging screen from a Setup page, select SCAN.
■ To go to the real-time imaging screen when the Setup menu is displayed and Scan is
highlighted, turn the small Rotary knob.
Probes
The Freestyle Probes page is a list of the most recently used probes by Name (type of probe),
Serial Number and date Last Used.
Locating the Active Probe
If a probe is currently turned on and wirelessly connected to the system, pressing the IDENTIFY
Soft Key causes the probe to beep. This can be useful when trying to locate a probe. To stop
the probe from beeping, press the IDENTIFY Soft Key a second time.
NOTE: For the Identify function to operate, the probe must be: on, linked to the system, and within
range of the Bluetooth radio.
Connecting and Disconnecting Probes
■ The CONNECT Soft Key connects the highlighted probe on the list if it is turned on and in a
three meter range of the console, or plugged into the system console using the probe Cable
Adapter.
■ The DISCONNECT Soft Key terminates the connection to the active probe.
For a description of the routine method of connecting probes, see the Scanning and
Wireless Probe Operation chapters of this Manual.