Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic
02.05 CS SD 24
Service 19
Page 19 of 20
Medical Solutions
5-
4Service
Functional checks 0
Check the function of the two keys to switch the laser light localizer on and off.
Switch the laser light localizer back on. The laser beams must go off automatically af-
ter approx. one minute.
Check the labels (see (Fig. 11 / p. 14)).