8 General information
ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG
02.05 CS SD 24
Page 8 of 20
Medical Solutions
The measuring procedure and the measuring device used (designation and serial num-
ber) are also to be documented.
NOTE
For evaluation purposes, the first measured value and the values
documented during maintenance or safety checks are to be com-
pared to the measured values. A sudden or unexpected increase
in the measured values may indicate a defect in the protective con-
ductor connections (protective conductor or contacts) - even if the
limit value of 0.2 ohms is not exceeded.
Laser light localizer option 0
(Comment: Information applies to line-type lasers of laser class 1M, e.g., laser light
localizers, I.I. / ARCADIS Avantic)
CAUTION
Laser emissions!
This product contains lasers of the class 1M. (USA: Laser class
1M)
Non-compliance can result in injury.
¹ Observe the safety information in ARTD-002.731.03...
When working with the laser light localizer, do not look
directly into the laser beam.
NOTE
Laser emissions!
There is no direct hazard to the eye (blinking reflex). Nevertheless
do not look directly into the laser beam.

Other manuals for Siemens ARCADIS Avantic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Avantic and is the answer not in the manual?

Siemens ARCADIS Avantic Specifications

General IconGeneral
BrandSiemens
ModelARCADIS Avantic
CategoryMedical Equipment
LanguageEnglish

Related product manuals