Siemens AG SPR2-330.814.02.01.02 ARCADIS Avantic
02.05 CS SD 24
General information 5
Page 5 of 20
Medical Solutions
Safety information 0
General safety information (in existing documents) 0
WARNING
Danger of injuries, death or material damage.
Non-compliance can lead to death, injury or material damage.
Please note:
¹ The product-specific safety notes in these instructions,
¹ The general safety information in TD00-000.860.01... and
¹ The safety information in accordance with ARTD Part 2.
WARNING
Electrical voltage!
Non-compliance can lead to severe bodily injury and even death.
¹ The internal uninterruptible power supply of the system
(UPS) provides several components with line voltage -
even when the system is switched off or the system
power plug is disconnected.
¹ Before the system is programmed to the local line volt-
age/line frequency, it is to be disconnected from the
power supply (the power plug is to be disconnected) and
the voltage output plug of the UPS is to be disconnected
as well.
Protective conductor resistance test information 0
Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
The protective conductor resistance of 0.2 ohms must not be exceeded.
First measured value
In the case of systems thoroughly tested at the factory prior to delivery, the protective con-
ductor resistance test was already performed and the measured values were recorded in
the system test report.
The measurements were performed using the measuring procedures and measuring
equipment recorded in the system test report.
The test report is included with the accompanying documentation.
If no covers were opened while installing the system and no additional components (e.g.
options) were installed or modified, the values recorded in the test report including the
measuring point (e.g. cover of the basic unit or cover of the image intensifier) can be used
in the protective conductor resistance report as the first measured value.

Other manuals for Siemens ARCADIS Avantic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Avantic and is the answer not in the manual?

Siemens ARCADIS Avantic Specifications

General IconGeneral
BrandSiemens
ModelARCADIS Avantic
CategoryMedical Equipment
LanguageEnglish

Related product manuals