6 General information
ARCADIS Avantic SPR2-330.814.02.01.02 Siemens AG
02.05 CS SD 24
Page 6 of 20
Medical Solutions
If the power plug was changed during system installation, covers were removed, or addi-
tional components were installed or modified, the values provided in the test report are
invalid.
The values are to be labeled invalid. Cross out the values, enter the comment "invalid
value" and confirm this with name, date and signature.
The protective conductor test must be performed again after all work has been completed.
In the case of systems with no measured values for the protective conductor test recorded
in the system test report, the protective conductor test is to be performed after all work
has been completed.
Measurement
The measurement must be performed according to DIN VDE 0751, Part 1 (see ARTD
Part 2). In this case the protective conductor resistance for all touchable conductive parts
must be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for a protective conductor connection.
During the measurement the power cable and additional connection cable with the inte-
grated protective conductor (e.g., monitor cable between the basic unit and monitor trol-
ley) must be moved section by section to detect cable breaks.
The protective conductor resistance must not exceed 0.2 ohms.
The values, including the measuring points, must be recorded as first measured values in
the protective conductor resistance report.
The measuring procedure and the measuring device used (designation and serial num-
ber) are also to be documented.
Separate the page with the report from these instructions and file it in the "Reports" regis-
ter in the system binder or log book.

Other manuals for Siemens ARCADIS Avantic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Avantic and is the answer not in the manual?

Siemens ARCADIS Avantic Specifications

General IconGeneral
BrandSiemens
ModelARCADIS Avantic
CategoryMedical Equipment
LanguageEnglish

Related product manuals