Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic
12.06 CS PS SP
General Startup 25
Page 25 of 66
Medical Solutions
Function check of the EMERGENCY STOP switches 2.1
• Connect the system to the voltage and boot it.
• Click "Emergency" for the patient registration.
• Activate the EMERGENCY STOP switch on the C-arm.
• The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
• Disengage the EMERGENCY STOP switch.
• The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
• Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop" (C-arm movement and emergency stop / p. 60).